2% calibration accuracy, ease of use and high availability combine to
achieve superior measurement confidence.
Detector & calibration options cover a wide range of connector
types, fiber types, common wavelengths and power levels from +24 to -60 dBm.
Features
• Shirt pocket size with spring clip
• Patented
low cost Interchangeable
connector
• Multi-fiber ID for fiber identification
• Large
sunlight readable display
• Displays dBm, dB, linear, tone Hz
• Power averaging mode for modulated signal
• Limited Feature mode for low skill measurement
• Simple to use
• 300-hour battery life
• Max / Min recording & display hold
• 9 calibrated wavelengths
• Compact, rugged & light weight
• 3-year calibration cycle
• 3-year
warranty
• Made in Australia
KI 9600 Series - Optical Power
Meter
The small KI 9600A Pocket Fiber Meter is ideal for measuring absolute /
relative light levels or test tones on single mode, multimode or plastic
optical fiber (POF) systems. High traceable accuracy and ease of use make it
perfect for field or laboratory.
Tough construction includes moisture resistance, rubber corners, a
captive connector dust cap and it can be dropped over 2 meters onto a hard
surface. This instrument meets MIL PRF 28800F Class 2.
When used with multiple KI 9800 sources, the Multi-Fiber ID feature
uniquely identifies up to 12 fibers.
The tight total uncertainty specification covers the entire range of
measurement, temperature, connectors and fiber types, without warm up or dark
current offset. Calibration is fully traceable.
Operational savings come from a 3-year warranty & re-calibration
cycle, 300 hours of battery life, and fast operation.
The meter displays mW, μW, nW, dB, dBm to 0.01 dB resolution. A separate
reference for each λ can be stored.
A Power Averaging Mode measures the average power of modulated signals.
A Limited Feature Mode enables a site manager to lock and track
instrument settings to reduce measurement skill, and improve both test
confidence and traceability.
Interchangeable optical connectors are dust and drop protected. Other
styles include the popular LC.
The InGaAs meter is the preferred solution for single mode testing from
900 – 1650 nm.
Ge meters offer modest accuracy from 660 to 1550 nm.
H series meters are available for high power testing. They offer good
immunity to wavelength and reflection effects.
For PON testing, the Wavelength Selective meter KI 9600WS01-Ge offers a
simple way to measure 1550 nm light only.
For testing 1 mm POF, ribbon fiber, MT-RJ, expanded beam connectors
etc., refer to the alternative KI 9600-XL brochure for instruments with large
area detectors.
Response
l
nm
|
Damage
level
dBm
|
Calibration l
nm
|
Power
range
dBm
|
Tone & mulita-fiber ID
sensitivity
dBm
|
Midrange
linearity1
dB
|
Calibration
Accuracy2
%
|
Polarization
Sensitivity
dB
|
Total
Uncertainty3, 5
dB
|
l
Sensitivity5
± 30 nm dB
|
InGaAs detector
|
600 ~ 1700
|
+15
|
660, 850
1300, 1310,
1390, 1490, 1550, 1610, 1625
|
+5 ~ -60
|
-40
-50
|
0.02
|
2 %
(0.09 dB)
|
< 0.005
|
0.3
|
0.03
|
H3B (InGaAs) detector
|
800 ~ 1700
|
+274
|
850
1300, 1310,
1390, 1490,
1550, 1590,
1610, 1625
|
+24 ~ -40
|
-20
-30
|
0.02
|
2 %
(0.09dB)
|
< 0.005
|
0.3
|
0.03
|
H5 (InGaAs) detector
|
800 ~ 1700
|
+25
|
850
1300, 1310,
1390, 1490, 1550, 1590, 1610, 1625
|
+15 ~ -50
|
-30
-40
|
0.02
|
2 %
(0.09dB)
|
< 0.005
|
0.3
|
0.03
|
Ge detector
|
600 ~ 1650
|
+15
|
635,650,660,780,1610, 1625
850, 1300,
1310, 1390, 1490, 1550
|
+10 ~ -60
|
-40
-50
|
0.04
|
2 %
(0.09dB)
|
< 0.005
|
0.5
|
0.04
|
|
|
|
|
|
typical
|
|
typical
|
max
|
typical
|
TECHNICAL
Specifications
Note 1: Midrange linearity excludes top 5 dB
and bottom 10 dB of range.
Note 2: Calibration condition: non
coherent light, -35±5 dBm, 23±1°C, ±1 nm, 10±3 nm FWHM, PC ceramic connector, 100 mm fiber.
Note
3: Includes contributions of: varying optical connector
types, calibration uncertainty, linearity over temperature & range, and
fiber core diameter up to 200 mm.
Note 4: H5 can sustain the damage level for
2 minutes.
Note 5: At calibration wavelengths in bold type.
Australian and international patents. Technical data is subject to
change without notice as part of our program of continuous improvements.
GENERAL Specifications
Parameters
|
Value
|
Battery life
Size
Weight
|
300 hours
124 x 81 x
25 mm, 4.9 x 3.2 x 1.0”
0.15 kg,
0.33 lb. Shipping 0.5 kg, 1.1 lb.
|
Operating / Storage
Relative humidity
Case
Tone detection
Calibration cycle
|
-15 to 55
°C / -25 to 70 °C
0 ~ 95%
Polycarbonate
with captive dust cap, 2.5-meter drop tested
200 ~ 2500
Hz ± 2 %
3 years
|
Max / min
Power
|
Recording
feature for stability testing
2 alkaline
AAA cells. Selectable auto-off, low battery indicator
|
ORDERING INFORMATION
Description
|
Part number
|
Instrument,
Power Meter InGaAs
|
KI 9600A-InGaAs
|
Instrument,
Power Meter H3B
|
KI 9600A-H3B
|
Instrument,
Power Meter H5
|
KI 9600A-H5
|
Instrument,
Power Meter Ge
|
KI 9600A-Ge
|
STANDARD
ACCESSORIES
Description
|
Quantity
|
Option, Hybrid Adaptor, Ceramic Sleeve, SC/SC
(OPT046)
|
1
|
Carry pouch (OPT156*)
|
1
|
Wrist strap
|
1
|
Calibration certificate
|
1
|
QA certificate
|
1
|
Quick guide
|
1
|
This instrument is supplied
with metal-free sleeve optical interchangeable connector adaptors. The power
meter works with both PC and APC connectors. The visible laser connector
ferrule type is fixed and customer specified as either PC or APC. Green is associated
with APC.
OPTIONAL
ACCESSORIES
Description
|
Part number
|
Option, accessories pack KI9000, includes:
1) Option,
Hybrid Adaptor, Ceramic Sleeve, SC/ST (OPT040)
2) Option,
Hybrid Adaptor, Ceramic Sleeve, SC/LC, metal body (OPT076)
|
OPT148
1
1
|
3) Soft carry
pouch
4) AAA
alkaline batteries
|
1
2
|
OPTIONAL
INTERCHANGEABLE CONNECTOR ADAPTORS
Description
|
Part number
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/FC
|
OPT051
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/LC, metal body
|
OPT076
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/ST
|
OPT040
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/D4
|
OPT055
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/MU
|
OPT080
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/LSA-DIN47256
|
OPT071
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/E2000
|
OPT060
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/E2000 Green
|
OPT060G
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/Universal 1.25 mm
|
OPT084
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/Universal 2.5 mm
|
OPT081
|
Option,
Hybrid Adaptor, Metal Sleeve, SC/SMA 905/906
|
OPT082
|
Option,
Hybrid Adaptor, Ceramic Sleeve, SC/F3000 or LC Simplex, plastic body
|
OPT072
|
The
power meter works with both PC and APC connectors.
History
Record
Revision
|
Date
|
Editor
|
Change Description
|
16
|
21/10/2016
|
TO Ng
|
Removed
specification section for KI9600WS01-Ge.
This
part number was also removed from theORDERING INFORMATION section.
Justification
for the change: see attachment….
|
|